Airvo™ 2 Nasal High Flow/HFNC System, Recertified

Airvo™ 2 Nasal High Flow/HFNC System, Recertified

Brand: Fisher & Paykal
Item Number: AIRVO2 R
Condition: DevicesMedical Devices
Log in to see your price List price: $2,995.00 (Each)
1

Description

The Fisher and Paykel Airvo™ 2 is a humidifier with integrated flow generator that delivers high flow warmed humidified respiratory gases to spontaneously breathing patients through a variety of patient interfaces. The Airvo™ 2 has an adjustable flow from 2 - 60 L/min depending on the patient interface. The Airvo™ HFNC is suitable for hospitals and long-term care facilities.

Specifications:

  • Dual spiral heated breathing tube with an integrated temperature sensor.
  • Can be used with Optiflow nasal cannula interfaces as well as direct-connect tracheotomy and mask interfaces.
  • Oxygen from a concentrator or cylinder can be added.
  • Built-in ultrasonic analyzer requires no calibration, service, or replacement.
  • Adjustable temperature and flow settings.
  • Designed for simple setup, use, and cleaning.

Fisher and Paykel Airvo™ 2 Specifications

Dimensions:

  • Height: 11.6” (29.5 cm)
  • Width: 6.7” (17 cm)
  • Depth: 6.9” (17.5 cm)
  • Weight: 4.8 lbs (2.2 kg)

General:

  • Supply frequency: 50-60 Hz
  • Supply voltage/current: 100-115 V 2.2 A (2.4 A max†); 220-240 V 1.8 A (2.0 A max)
  • Sound pressure level: Alarms exceed 45dbA @ 1 m
  • Auditory alarm pause: 115 seconds
  • Expected service life: 5 years
  • Serial port: The serial port is used for downloading product data, using F&P Infosmart software.
  • Warm-up time: 10 minutes to 31 °C (88 °F), 30 minutes to 37 °C (98.6 °F) using a
  • MR290 chamber with flow rate of 35 L/min and starting temperatur e 23 ± 2 °C (73 ± 3 °F)
  • Target temperature settings: 37, 34, 31 °C

Performance:

  • Humidity performance
    • >33 mg/L at 37 °C target
    • >12 mg/L at 34 °C target
    • >12 mg/L at 31 °C target
  • Maximum temperature of delivered gas: 43 °C (109 °F) (in accordance with ISO 80601-2-74)
  • Maximum surface temperature of applied parts: 44 °C (111 °F) (in accordance with ISO 80601-2-74)
  • Flow range (default): 10-60 L/min
  • Flow range (Junior Mode): 2-25 L/min
  • Maximum oxygen input: 60 L/min
  • Oxygen analyzer accuracy: < ± 4 % (within the range 25-95% O2)
  • Operating conditions: 18-28 °C (64-82 °F), 30-70% RH

REFERENCE LIBRARY

Product Brochure

Quick Setup Guide

Disinfection Kit Manual

WHAT’S INCLUDED

  • AC Power Cord
  • Quick Setup Guide *pdf download
  • 12-Month Warranty on Parts & Labor